The display limits for the different units of measurement are summarized in the table
below.
The lower display limit may be increased by the user. Please refer to Section
for how to edit the lower display limit.
As soon as an acoustic alarm sounds, the tip should be removed from the spot being
tested. After displaying a constant leak rate the ZERO button should be pressed
again to repeat the test. Thus, a measurement error can be prevented and the leak
can be located.
Fig. 3-6 Sniffer display when detecting a leak
Unit
Lower display limit
Upper display limit
g/a
0.05
1,000
oz/yr
0.002
1,000
ppm
0.5
100,000
mbar l/s
1x10
-7
9.9x10
-2
Pa m³/s
1x10
-8
9.9x10
-3
atm cc/s
1x10
-7
9.9x10
-2
Torr l/s
1x10
-7
9.9x10
-2
Содержание Ecotec E3000
Страница 2: ...0 2 kina22eIVZ fm Technical Handbook 1206 ...
Страница 13: ...General Information 1 3 kina22e chapter 1 fm technical handbook 1206 ...
Страница 30: ...1 20 General Information kina22e chapter 1 fm technical handbook 1206 ...
Страница 38: ...2 8 How the Instrument Works kina22e chapter 2 fm technical handbook 1206 ...
Страница 58: ...3 20 Operation of the Ecotec E3000 kina22e chapter 3 fm technical handbook 1206 ...
Страница 98: ...4 40 Equipment Settings kina22e chapter 4 fm technical handbook 1206 ...
Страница 106: ...6 8 Ecotec E3000 Messages kina22e chapter 6 fm technical handbook 1206 ...
Страница 133: ...Maintenance 8 18 kina22e chapter 8 fm technical handbook 1206 ...
Страница 143: ...Annex 9 10 kina22e chapter 9 fm technical handbook 1206 9 2 Declaration of Conformity ...