6)
E38
Capillary broken (>60 s)
Actual flow is higher than upper flow
limits for more than 60 s, pumps will be
switched off to protect filaments
Capillary broken or leaky
Upper pressure limit setting too low
Replace sniffer line and recalibrate!
Or: Replace probe cable* and
recalibrate!
Increase upper flow limit (see
Section
)
Press OK to restart the Ecotec
E3000
E39
No emission!
Start emission failed on both filaments
during start-up!
After extended shutsown times this
error may occur during the first
10 min after power on
Please acknowledge the error and
startup again automatically.
If problem persists, call nearest
INFICON service representative.
E40
No emission!
Emission failed during operation
After extended shutdown times this
error may occur during the first 10
min after power on.
Please acknowledge the error and
startup again automatically.
If problem persists, call nearest
INFICON service representative.
E41
No communication with Transpector!
Software cannot establish
communication with Transpector
Check cable connection between
Transpector and CPU-board!*
Call nearest INFICON service
representative!
E42
Transpector temperature > 70°C or <
0°C!
Air filter dirty
Ambient temperature too high
Ambient temperature too low
Replace main air filter (see Section
Check ambient temperature
E43
Transpector limit exceeded!
Refer to Transpector internal data
Call nearest INFICON service
representative!
E44
Transpector communication error!
Refer to Transpector internal data
Call nearest INFICON service
representative!
E45
Transpector hardware error!
Refer to Transpector internal data
Call nearest INFICON service
representative!
E46
Transpector hardware warning!
Refer to Transpector internal data
Call nearest INFICON service
representative!
E47
Transpector over-pressure!
After extended shutdown times this error
may occur during the first 10 min after
power on.
This may also occur if sniffer line has
been disconnected for extended times!
Please acknowledge the error and
the Ecotec E3000 will startup again
automatically.
Reconnect sniffer line and
acknowledge error, Ecotec E3000
will startup again automatically.
E48
Transpector no emission!
After extended shutdown times this error
may occur during the first 10 min after
power on.
This may also occur if sniffer line has
been disconnected for extended times!
Please acknowledge the error and
the Ecotec E3000 will startup again
automatically.
Reconnect sniffer line and
acknowledge error, Ecotec 3000 will
startup again automatically.
Error no.
Plain text message
Possible reason
Recommendation
Содержание Ecotec E3000
Страница 2: ...0 2 kina22eIVZ fm Technical Handbook 1206 ...
Страница 13: ...General Information 1 3 kina22e chapter 1 fm technical handbook 1206 ...
Страница 30: ...1 20 General Information kina22e chapter 1 fm technical handbook 1206 ...
Страница 38: ...2 8 How the Instrument Works kina22e chapter 2 fm technical handbook 1206 ...
Страница 58: ...3 20 Operation of the Ecotec E3000 kina22e chapter 3 fm technical handbook 1206 ...
Страница 98: ...4 40 Equipment Settings kina22e chapter 4 fm technical handbook 1206 ...
Страница 106: ...6 8 Ecotec E3000 Messages kina22e chapter 6 fm technical handbook 1206 ...
Страница 133: ...Maintenance 8 18 kina22e chapter 8 fm technical handbook 1206 ...
Страница 143: ...Annex 9 10 kina22e chapter 9 fm technical handbook 1206 9 2 Declaration of Conformity ...