Gas select, gas a, gas b
Through the control inputs
GAS
(pin 9, 13 and 25 of the 25 way Sub-D connector) it
is possible to select a gas loaded in the active gas memory (one gas out of four). The
Ecotec E3000 will then operate in the single gas mode.
With an edge triggered signal from
LOW
to
HIGH
at Pin 25, the selected gas at Pin 13
and 9 becomes valid.
7.1.4
PLC Outputs
These outputs can be used to monitor the Ecotec E3000 via a programmable logic
control (PLC).
PLC Outputs are designed as an „Open Collector Output“ or a relay output. Please
refer to the example given below for an open collector output „Error“.
Technical data
Pin 25
Pin 13
Pin 9
Function
0
x
x
Gas selecting disabled
1
0
0
1
st
gas is selected
1
0
1
2
nd
gas is selected
1
1
0
3
rd
gas is selected
1
1
1
4
th
gas is selected
Fig. 7-2 Circuit for open collector output “Error”
6
17
24 V
I/O Port (open collector output)
Error active = 0 Volt
Error inactive = 24 Volt
10 K
Caution
Permissible max. voltage and current for open collector outputs are: 28 V; 50 mA.
Содержание Ecotec E3000
Страница 2: ...0 2 kina22eIVZ fm Technical Handbook 1206 ...
Страница 13: ...General Information 1 3 kina22e chapter 1 fm technical handbook 1206 ...
Страница 30: ...1 20 General Information kina22e chapter 1 fm technical handbook 1206 ...
Страница 38: ...2 8 How the Instrument Works kina22e chapter 2 fm technical handbook 1206 ...
Страница 58: ...3 20 Operation of the Ecotec E3000 kina22e chapter 3 fm technical handbook 1206 ...
Страница 98: ...4 40 Equipment Settings kina22e chapter 4 fm technical handbook 1206 ...
Страница 106: ...6 8 Ecotec E3000 Messages kina22e chapter 6 fm technical handbook 1206 ...
Страница 133: ...Maintenance 8 18 kina22e chapter 8 fm technical handbook 1206 ...
Страница 143: ...Annex 9 10 kina22e chapter 9 fm technical handbook 1206 9 2 Declaration of Conformity ...