Equipment Settings
4-15
ki
n
a
22
e c
h
ap
te
r 4
.fm
te
chn
ica
l ha
nd
boo
k
(1
20
6)
Audio intern
The internal loudspeaker of the main unit can be switched on and off. This will not
effect the headphones outlet.
Default value: enabled.
Audio sniffer
The alarm of the loudspeaker in the probe handle can be set to react to exceeding
the
TRIGGER
LEVEL
or exceeding the
SEARCH
LEVEL
. It can also be
DISABLED
.
Default value:
TRIGGER
LEVEL
.
Alarm type
Different types of audio alarm of the main unit speaker can be selected. Options are
SETPOINT
,
PINPOINT
, and
TRIGGER
ALARM
Default value: Trigger Alarm
Default value: Trigger Alarm
Volume
The
MINIMUM
VOLUME
and the currently selected
VOLUME
for the loudspeaker in the
main unit can be edited.
The
MINIMUM
VOLUME
is the lowest level the volume can be set to from the main
display and should prevent unintentional turning off of the main unit audio alarms.
Both values will also be valid for the headphones outlet.
Both volumes (minimum and current) can be chosen between 0 and 15.
Default value for minimum volume: 2
Default value for the current volume: 2
SETPOINT
The frequency of the tone will change when the trigger level is
exceeded. Below the trigger level the loudspeaker will remain off.
TRIGGER
/
ALARM
An acoustic alarm with a continuous low tone will sound as soon as
the search level is exceeded. An acoustic alarm which consists of
two tones sounds as soon as the trigger level is exceeded. Below the
search level the loudspeaker will remain off.
If the Ecotec E3000 is set to
TRIGGER
ALARM
as the alarm type an
additional button will be shown on the lower right side called
TONE
1,2,3. With this button different alarm sounds can be selected for the
trigger alarm to clearly distinguish between nearby other Ecotec
E3000 leak detectors.
PINPOINT
The external loudspeaker is always on. Within the leak rate range of
+/- one decade of the trigger level the frequency will change
depending on the leak rate. Outside this range the tone will be
constantly low or high. This setting is recommended if also leaks
which are still below the trigger level need to be pinpointed or if the
expected leak rates are within the range of the trigger level.
Содержание Ecotec E3000
Страница 2: ...0 2 kina22eIVZ fm Technical Handbook 1206 ...
Страница 13: ...General Information 1 3 kina22e chapter 1 fm technical handbook 1206 ...
Страница 30: ...1 20 General Information kina22e chapter 1 fm technical handbook 1206 ...
Страница 38: ...2 8 How the Instrument Works kina22e chapter 2 fm technical handbook 1206 ...
Страница 58: ...3 20 Operation of the Ecotec E3000 kina22e chapter 3 fm technical handbook 1206 ...
Страница 98: ...4 40 Equipment Settings kina22e chapter 4 fm technical handbook 1206 ...
Страница 106: ...6 8 Ecotec E3000 Messages kina22e chapter 6 fm technical handbook 1206 ...
Страница 133: ...Maintenance 8 18 kina22e chapter 8 fm technical handbook 1206 ...
Страница 143: ...Annex 9 10 kina22e chapter 9 fm technical handbook 1206 9 2 Declaration of Conformity ...