8-13
Maintenance
ki
n
a
22
e c
h
ap
te
r 8
.fm
te
chn
ica
l ha
nd
boo
k
(1
20
6)
8.7.1
Switching from metal to plastic capillary filter
When you switch from the metal capillary filter to the plastic capillary filter you need
to remove the taper gasket. The plastic capillary filter will not fit with the taper gasket
installed.
1
Remove the two Philips screws in the flange of the sniffer tip and remove sniffer tip.
2
Take a small pin or needle (about 0.5 mm) and push the steel capillary out off the tip
from the top.
Fig. 8-14 Pushing out the steel capillary
Fig. 8-15 Capillary sticking out of sniffer tip flange
Содержание Ecotec E3000
Страница 2: ...0 2 kina22eIVZ fm Technical Handbook 1206 ...
Страница 13: ...General Information 1 3 kina22e chapter 1 fm technical handbook 1206 ...
Страница 30: ...1 20 General Information kina22e chapter 1 fm technical handbook 1206 ...
Страница 38: ...2 8 How the Instrument Works kina22e chapter 2 fm technical handbook 1206 ...
Страница 58: ...3 20 Operation of the Ecotec E3000 kina22e chapter 3 fm technical handbook 1206 ...
Страница 98: ...4 40 Equipment Settings kina22e chapter 4 fm technical handbook 1206 ...
Страница 106: ...6 8 Ecotec E3000 Messages kina22e chapter 6 fm technical handbook 1206 ...
Страница 133: ...Maintenance 8 18 kina22e chapter 8 fm technical handbook 1206 ...
Страница 143: ...Annex 9 10 kina22e chapter 9 fm technical handbook 1206 9 2 Declaration of Conformity ...