Monitoring Sensitivity
It is highly recommended to always keep the
SENSITIVITY
CHECK
function enabled.
This function should only be disabled when operating in argon-free environment as
the argon signal is used for monitoring.
For details on this function see Section
(“Sensitivity Check”)
If sensitivity changes,
ERROR
CODE
E
30 will be displayed. In this case, please
recalibrate the Ecotec E3000 to restore sensitivity (see Section
).
The error code will be displayed every 15 seconds until a recalibration is started.