Default setting: logarithmic
Recorder output gas
Here one gas from the four gases is defined which will be output as the analog signal
through channel 1. This is selected through the
PLUS
/
MINUS
keys.
Selecting one of the numbers 1 ... 4 outputs the corresponding gas 1 ... 4 on the
display as the analog signal.
Channel 2 (pin 14 of the I/O port) indicates the gas number by way of a voltage, i.e.
a voltage between 1 V ... 4 V will be present accordingly.
When selecting
AUTO
the leak rate which has the highest value with reference to the
corresponding trigger level is output as the analog signal.
Correspondingly the voltage in channel 2 will differ accordingly.
In the modes "
ERROR
”, "
NOT
READY
TO
MEASURE
” and "
SLEEP
” a voltage of U = 10 V
will be output through channels 1 and 2.
Default setting: AUTO
Trigger
Leak rate
Output voltage
0.1 oz/yr
1.5 oz/yr
5.35V
3 g/a
20 g/a
5.60V
50 g/a
6.39V
5 x 10
-4
mbar l/s
8 x 10
-5
mbar l/s
2.806V
6 x 10
-3
mbar l/s
6.556V
Содержание Ecotec E3000
Страница 2: ...0 2 kina22eIVZ fm Technical Handbook 1206 ...
Страница 13: ...General Information 1 3 kina22e chapter 1 fm technical handbook 1206 ...
Страница 30: ...1 20 General Information kina22e chapter 1 fm technical handbook 1206 ...
Страница 38: ...2 8 How the Instrument Works kina22e chapter 2 fm technical handbook 1206 ...
Страница 58: ...3 20 Operation of the Ecotec E3000 kina22e chapter 3 fm technical handbook 1206 ...
Страница 98: ...4 40 Equipment Settings kina22e chapter 4 fm technical handbook 1206 ...
Страница 106: ...6 8 Ecotec E3000 Messages kina22e chapter 6 fm technical handbook 1206 ...
Страница 133: ...Maintenance 8 18 kina22e chapter 8 fm technical handbook 1206 ...
Страница 143: ...Annex 9 10 kina22e chapter 9 fm technical handbook 1206 9 2 Declaration of Conformity ...