0-10
Important Safety Precautions
ki
n
a
22
eI
V
Z
.fm
Tech
ni
ca
l Ha
nd
bo
ok
(1
20
6)
Caution
The air filter should be checked for contamination at least every 6 months and
should be definitely exchanged after 2 years.
Caution
The lubricant reservoir may contain toxic substances from the pumped media.
Please dispose of lubricant reservoir as required by local regulations. A Safety Data
sheet for the lubricant is available on request.
Caution
Caution: Abrupt movements.
Abrupt movements can damage the running turbo pump.
Avoid abrupt movement and vibration of the instrument (e.g. running over cables,
door sills) during operation and up to 4 minutes after switching off since the turbo
pump can be damaged.
Содержание Ecotec E3000
Страница 2: ...0 2 kina22eIVZ fm Technical Handbook 1206 ...
Страница 13: ...General Information 1 3 kina22e chapter 1 fm technical handbook 1206 ...
Страница 30: ...1 20 General Information kina22e chapter 1 fm technical handbook 1206 ...
Страница 38: ...2 8 How the Instrument Works kina22e chapter 2 fm technical handbook 1206 ...
Страница 58: ...3 20 Operation of the Ecotec E3000 kina22e chapter 3 fm technical handbook 1206 ...
Страница 98: ...4 40 Equipment Settings kina22e chapter 4 fm technical handbook 1206 ...
Страница 106: ...6 8 Ecotec E3000 Messages kina22e chapter 6 fm technical handbook 1206 ...
Страница 133: ...Maintenance 8 18 kina22e chapter 8 fm technical handbook 1206 ...
Страница 143: ...Annex 9 10 kina22e chapter 9 fm technical handbook 1206 9 2 Declaration of Conformity ...