Use these mechanical and electrical wiring specifications.
Mechanical:
o
Solid Wire: 18-12 AWG
o
Stranded Wire: 18-12 AWG
o
Torque: 4.0 lb in (0.45 N m)
Electrical: Shield wiring, or conduit, at the source
8
Connect optional Ethernet cable
If an Ethernet connection will be used, connect the cable to its port.
9
Secure the cable and power line
Use the guides to secure the already connected Ethernet cable and
power line.
•
Lift any guide by its tab.
•
Place the cable or power line in the guide's groove.
•
Lower and secure the tab to the guide base.
10
Close case lid and secure to case base
Ensure all wires and cables are:
•
secure
•
contained entirely inside the case
•
away from the unit's closure surfaces
This supports the integrity of the items and full closure of the lid. If the
strain relief on the Ethernet cable resists full closure of the unit, use a 90°
adapter or short adapter cable. Close the case lid from left to right.
Using a T25 torx screwdriver, reinstall the four fasteners that secure the
case lid to the case base. Tighten them to a torque value of 1.69 N m
(240 oz in).
Figure 4.3 Installation that
includes
an external power source, Ethernet, or both
Operation Precheck
Power on the unit: On the case lid, press the power button for approximately 5 seconds until the power-
button light turns on.
Observe the color and pattern of the power-button light, which indicates the unit's operational status. If the
light is solid green (not blinking)* the unit is fully operational. For information about other indicator-light
patterns, refer to
Table 3.6 Power-button light indicators
*
Note
: When an RGX Gateway operating in LENS repeater mode
is fully operational its power-button light displays a very
slowly
blinking green light.