32
ISERTPS
a
RGX Gateway
b
5 m (16 '
4 ″)
1 m (3 ' 3″)
Ground
Figure 4.1 Distance requirements: RGX-to-power-supply and RGX-to-ground
a
ISERTPS shown, also applies to ERTPS.
b
Also applies to
an RGX Gateway operating in LENS repeater mode
.
Placement and Mounting
The unit can be placed or mounted for operation.
The optional Wall Mount Kit can be used for a unit that will be mounted to a wall, another surface, or to a
strut-type rail setup. The kit contains a label that replicates important information printed on the back of the
RGX, including the LAN MAC and WAN MAC values. Copy these two values onto the kit label in the
designated spaces. Affix the label to an area of the unit
that will be visible during operation.
The optional Magnet Mount Kit is used to mount the unit to a metal surface.
IMPORTANT:
Do not
use the
Magnet Mount Kit with units that draw operational power from a fixed DC power source.
Installation Preparation
Based on how the RGX will be mounted (or placed) and its intended operational power source and
communications options, use the information below to determine which preparation tasks are needed.
Refer to the
section of this manual for task instructions.
Table 4.2 Possible installation preparation tasks
Application
Service task
Mounting
Wall, surface, strut-type rail mounting
Wall-mount installation
Magnet mounting
Magnet-mount installation;
do not use magnet mounts
with units that
will draw operational power from a fixed DC power source
Operational power source
Compatible power-supply accessory from Industrial
Scientific
IS Cable Adapter installation
Fixed DC power source
Conduit-hub plug removal