C O M M A N D R E F E R E N C E
LDC-3908 Mainframe Command Reference
03_02
LDC-3908
37
C H A P T E R
4
!
The TERM command allows the programmer to change the default (IEEE488.2 standard)
terminator to include the carriage return.
Parameters
An <nrf value>, 0 = FALSE, non-zero = TRUE
Notes
An altered terminator will be in the form <CR><NL><^END>. Technically this takes the LDC-
3908 Laser Diode Controller out of IEEE488.2 specification, but this command may be used for
convenience when using non-standard GPIB controllers. This termination will be sent with all
output until the “TERM 0" command is sent or the LDC-3908 Laser Diode Controller is powered off.
Examples
“Term 1" -temporarily sets <CR><NL><^END> as the output terminator.
The TERM? query allows the programmer to determine whether the default (IEEE488.2 standard)
terminator has been altered to include a carriage return.
Parameters
None. The response is an <nrf value>, 1 = TRUE, 0 = FALSE.
Notes
An altered terminator will be in the form <CR><NL><^END>. This termination will be sent with
all output until the “TERM 0" command is sent or the LDC-3908 Laser Diode Controller is powered
off.
Examples
“Term?” -response: 1, means that the <CR><NL><^END> terminator is temporarily being
used.
"Term?" -response: 0, means that the <NL><^END> terminator (IEEE 488.2 standard) is being
used.
The TIME? query allows the programmer to determine how much time has passed since the LDC-
3908 Laser Diode Controller was last powered up.
Parameters
None. The response is character data in the form: hours:minutes:seconds.
Notes
The TIME clock is independent of the TIMER clock.
Examples
“Time?” -response: 00:01:02.36, means that 1 minute, 2.36 seconds have passed since the LDC-
3908 Laser Diode Controller was powered up.
"TIME?" -response: 00:32:00.76, means that 32 minutes, 0.76 second have passed since the LDC-
3908 Laser Diode Controller was powered up.
The TIMER? query allows the programmer to determine how much time has passed since the last
TIMER? query was issued.
Parameters
None. The response is character data in the form: hours:minutes:seconds.
Notes
Each time the TIMER? query is issued, the timer is reset to 0, and the elapsed time since the last
TIMER? query is returned.
Examples
“Timer?” -response: 00:02:00.31, means the LDC-3908 Laser Diode Controller has been on for 2
minutes, 0.31 seconds since the last TIMER? query was issued.
“TIMER?” -response: 00:00:12.03, means the LDC-3908 Laser Diode Controller has been on for
12.03 seconds since the last TIMER? query was issued.
TERM
F
RONT
P
ANEL
R
EMOTE
TERM?
F
RONT
P
ANEL
R
EMOTE
TIME?
F
RONT
P
ANEL
R
EMOTE
TIMER?
F
RONT
P
ANEL
R
EMOTE
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Содержание LDC-3908
Страница 3: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 7: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 9: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 11: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 53: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...