![ILX Lightwave LDC-3908 Скачать руководство пользователя страница 44](http://html1.mh-extra.com/html/ilx-lightwave/ldc-3908/ldc-3908_user-manual_2044768044.webp)
C O M M A N D R E F E R E N C E
LDC-3908 Mainframe Command Reference
03_02
LDC-3908
29
C H A P T E R
4
!
LDC-3908 Mainframe Command Reference
This section presents the mainframe device-dependent commands for both local and remote
operation of the LDC-3908 Laser Diode Controller, listed in alphabetical order. Also included in this
section are the descriptions of the generic module commands, see Figure 3.1. Refer to the
appropriate module instruction manual for details on each specific module. Figure 4.1 below
provides an example of the command description format used in the listings. Sub-sections for each
path are presented, listing the commands which are legal for that path. Those commands which
also emulate local (front panel) operation are denoted by the solid box next to the Local label in the
upper left corner of the command description.
Figure 4.1
Command Description Format
Command
Required letters are shown in
upper case; lower case letters
are optional.
Front Panel/
A solid box means the command
can be executed in that mode.
Command mnemonics in front
panel (local) mode may vary.
Command Description
Syntax Diagram
(For Remote Commands)
Parameters
Tells what parameters are
expected, and their type.
An <nrf value> refers to a
flexible numeric notation
(IEEE-488.2). Integer,
exponential, or floating point
may be used with the decimal
radix. Alternate radices may be
selected in remote operation.
Points of Interest
Has special information
about the command.
Examples
Shows typical statements
using the command.
Front Panel
Remote
SCRoll
feature of the status menu.
SYNTAX DIAGRAM
<white
space>
< nrf
value >
PARAMETERS
An <nrf value> where 1 = on, and 0 = off.
POINTS OF
INTEREST
The auto-scrolling feature does not persist after
the LDC-3908 is shut off.
The auto-scrolling rate is not adjustable.
The Status Menu shows only the channels which are
present in the LDC-3908.
EXAMPLES
Remote Execution
"SCR 1" - action: the auto-scrolling feature is
enabled.
"Scroll 0" - action: the auto-scrolling feature is
disabled.
The SCRoll command enables/disables the auto-scrolling
SCR
:
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Содержание LDC-3908
Страница 3: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 7: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 9: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 11: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 53: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...