I N T R O D U C T I O N A N D S P E C I F I C A T I O N S
Initial Inspection
2
LDC-3908
C H A P T E R
1
!
Initial Inspection
When you receive your LDC-3908 Controller, verify that the following items were shipped with the
instrument:
•
LDC-3908 Laser Diode Controller
•
LDC-3908 Mainframe Instruction Manual (optional)
•
Power Cord
Note:
When unpacking the instrument, be sure to save the packaging and the protective
ESD caps on the rear of the instrument in case you have to return the instrument to ILX
Lightwave or ship it elsewhere. Shipping damage is not covered under the standard
instrument warranty.
Installing Your LDC-3908 Laser Diode Controller
Grounding Requirements
The LDC-3908 Laser Diode Controller comes with a three-conductor AC power cable. The power
cable must either be plugged into an approved three-contact electrical outlet or used with a three-
contact or two-contact adaptor with the grounding wire connected to an electrical ground (safety
ground). The LDC-3908's power jack and supplied power cable meet IEC safety standards.
AC Line Power Requirements
You can operate the LDC-3908 Controller from a single-phase power source delivering nominal line
voltages of 100, 120, 220, or 240 VAC (all values RMS), at 48 to 66 Hz. The line power voltage can
vary -10% to +6%, but cannot exceed 250 VAC. Maximum power consumption is 1750 Volt-Amps
with 8 modules (LDC-3916376 1.5A/9W Controller modules). The instrument's operational voltage
setting is factory configurable and need not be changed before operating the instrument, unless it
is moved to a region with a different voltage.
Before connecting the LDC-3908 Laser Diode Controller to a power source, verify that
your facilities AC power source matches the setting of the LDC-3908's voltage printed on
the rear panel of the instrument.
To avoid electrical shock hazard, connect the instrument only to properly earth-
grounded, three-prong receptacles. Failure to observe this precaution can result in
severe injury or death.
RS-232 Connectors
The RS-232 interface consists of two 9-pin, D-sub connectors, located on the rear panel. The A
connector is used to connect to the host computer with a standard straight-through cable. See
Figure 1.2 -- LDC-3908 Rear Panel. The B connector is not enabled at this time.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Содержание LDC-3908
Страница 3: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 7: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 9: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 11: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 53: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...