O P E R A T I N G I N R E M O T E C O N T R O L
Remote Communication
20
LDC-3908
C H A P T E R
3
!
RS-232 Configuration
Before you can operate the LDC-3908 over RS-232, you need to know its baud rate. The
instrument's baud rate setting must match the baud rate used by the serial RS-232 interface, which
is typically a serial Comm port on your PC. The following sections describe reading and changing
the baud rate of the LDC-3908.
The LDC-3908's RS-232 is configured for 8-bit, no parity. This is not adjustable. Only the baud rate
may be altered.
Reading the Baud Rate
1
Press the Main switch on the front panel to recall the Main Menu.
2
Press the Sys. Config. soft-key switch, to reach the System Config. menu.
3
Press the Comm. soft key switch to access the Config Comm. menu.
The instrument will display the baud rate there.
Changing the Baud Rate
The factory default baud rate is 19200. If it is necessary to change the baud rate, select the Baud
Rate field of the Config. Comm. menu. Turn the Adjust Knob until the desired baud rate value is
displayed, and press the enter key. The new baud rate will then be stored in non-volatile memory.
The allowable baud rates are 1200, 2400, 4800, 9600, and 19200. Refer to the serial port
configuration of your PC (if used) for the appropriate baud rate.
Remote Communication
The following sections describe some of the issues regarding Remote operation. Before remote
communications are initiated, be sure to configure the GPIB or RS-232 as described above.
Changing Operation from Local to Remote
Sending a command over the GPIB or RS-232 interface will automatically put the instrument in
Remote mode. The Remote indicator on the lower right of the front panel will illuminate when the
controller is in remote operation mode. When the instrument is in Remote mode, pressing the Local
switch on the front panel returns the instrument to Local control mode unless the Local Lockout
state has been activated by the host computer (GPIB only). Local Lockout disables all LDC-3908
front panel switches until this condition is changed by the host computer. Front panel menu
changes are allowed by using the soft keys while in Remote mode. Changing channels is allowed
by using the knob or INC/DEC keys. Other items in menus such as setpoint adjustments, mode
changes, cursor movements, or turning the TEC/laser on or off are not allowed while in Remote
mode.
The Talk/Listen
indicator on the front panel is illuminated when the instrument is communicating
over the GPIB or RS-232 interface.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Содержание LDC-3908
Страница 3: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 7: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 9: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 11: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 53: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...