C O M M A N D R E F E R E N C E
LDC-3908 Mainframe Command Reference
03_02
LDC-3908
33
C H A P T E R
4
!
The CHECKSUM? query returns an integer value which is the summation of all pertinent
parameters of the present configuration.
Parameters
None. The response consists of an integer value.
Notes
The response value may be stored and compared at a later time to verify that the present
configuration of the instrument is the same as expected. This feature works well with the *RCL
(recall) command as a means of verifying a stored setup.
The checksum does not test the functionality of the instrument or which modules are present.
Examples
“CHECKSUM?” -response: 3812, is the checksum of the present configuration of parameters on
the instrument.
The DELAY command causes the execution of commands to be delayed by a user-defined time
interval. This command is similar to the *WAI common command, except that execution resumes
after the specified number of milliseconds, instead of waiting for the Operation-Complete flag to be
set.
Parameters
An <nrf value> which represents the delay time, in milliseconds.
Notes
Further commands are held off until the delay period elapses.
This command is useful for creating delays which do not require very much program code and do
not tie up the GPIB during execution.
Examples
“DELAY 500" -action: Further commands and queries are not executed until 0.5 second has
elapsed from the time this command is executed.
"Tec:T 22;Delay 2000;Tec:T?" -actions: The TEC is set to 22.0×C, then the LDC-3908 Laser Diode
Controller waits for about 2.0 seconds before measuring and returning the temperature.
The “ERRors?” query returns a list of “mainframe” device and command errors, followed by a
binary representation of the occurrence of “module” device and command errors. The “mainframe”
errors are those which have occurred since the last “ERR?” query. The “module” errors are not
cleared until the “MODERR?” query is sent to each module which reports an error. The errors are
notated by a number (code) which corresponds to the type of error which occurred. See Chapter 3,
Error Messages, for information regarding error handling.
Parameters
None. The response consists of one or more “mainframe” error code values, separated by
commas and a 8-bit binary representation of errors which exist on each channel.
Notes
A response of 0,00000000 indicates that no errors were reported. The first zero indicates no
“mainframe” errors, and the string of eight zeroes after the comma indicates no “module” errors.
The response data is sent as character data.
Examples
“ERR?” -response: 0,00000001 means no “mainframe” errors reported, but channel #1 has an
error. Use “Chan 1;MODERR?” to retrieve the error code(s) from channel 1.
"Errors?" -response: 219,11110000, means that the LDC-3908 was unable to restore the setup
parameters (at power-up) because of a change in modules (error 219). Also, there are errors
present on channels 5-8.
CHECKSUM?
F
RONT
P
ANEL
R
EMOTE
DELAY
F
RONT
P
ANEL
R
EMOTE
ERRors?
F
RONT
P
ANEL
R
EMOTE
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Содержание LDC-3908
Страница 3: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 7: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 9: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 11: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 53: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...