22
1693 RLC Digibridge
Operation
3.2.7 7000-05 Tweezers
The 7000-05 tweezers combine the two functions:
• They handle chip components, up to 0.5 in.
(12 mm) thick.
• Serve as a test fixture for measuring these com-
ponents, if their terminals are on opposite faces.
The 7000-05 Chip Component Tweezers can handle
small unleaded “chips” or SMDs (surface mounted
devices), passive sensors, and components for test-
ing on the Digibridge. A four-terminal Kelvin con-
nection extends to the tip of the tweezers where the
measurement becomes two-terminal, therefore series
impedance of the connecting cables and internal
tweezer connections do not affect the measurement.
The small amounts of residual-tip resistance and
inductance can be automatically corrected by using
the Digibridge shorting function with the tips pressed
together. Guard shields between the tweezer blades
minimize capacitance between them. A correction for
this small capacitance can be made using Digibridge
open circuit test with the tips held at a spacing equal
to that of the component’s contact spacing.
Figure 3-13: 7000-05 Chip Component
Tweezers
For this accessory an additional accuracy (for param-
eters listed) must be added to the standard instrument
accuracy as defined by AutoAcc.
C: ± 1.5 pF
R: ± 10 mΩ
L: ± 100 nH
Q: for R > 100 Ω, ±(9 * Freq * R * 10
-12
ppm)
Q; for R ≤ 100 Ω, ±((R/Freq) * 10
-6
ppm)
Measurement accuracy is very sensitive to connection
geometry. The same connection geometry must be
used for open/short compensation as for connection
to the device under test.
Connection to Digibridge:
Connect to
Digibridge
Cable Marking
Connection to DUT
PH (potential, high)
HS
to positive (+) terminal
of DUT
IH (current, high)
HD/HF
PL (potential, low)
LS
to negative (-) terminal
of DUT
IL (current, low)
LD/LF
The PH (HS) and IH (HD/HF) cables connect to the
fixed arm of the tweezers (+) and the PL (LS) and
IL (LD/LF) cables connect to the movable arm of
the tweezers (-). To ensure valid measurements it is
especially important to observe the correct polarity
when DC bias is to be used.
If dc bias is used (see paragraph 3.7), note that the
fixed arm of the tweezers is positive and the movable
arm is negative.
Содержание RLC Digibridge 1693
Страница 12: ...x Table of Contents 1693 RLC Digibridge This page is intentionally left blank...
Страница 24: ...xxii 1693 RLC Digibridge Abbreviated Specifications This page is intentionally left blank...
Страница 122: ...98 1693 RLC Digibridge Operation Table 3 28 Truth table of compacted binary format for output data...
Страница 130: ...106 1693 RLC Digibridge Theory Figure 4 1 Block diagram of the 1693 RLC Digibridge...
Страница 163: ...139 1693 RLC Digibridge Service and Maintenance Table 5 11 Entry of limits Table 5 12 Bin assignment check...
Страница 171: ......