114
1693 RLC Digibridge
Service and Maintenance
might flicker between 2 and 4. If, for example, ‘X” is
flickering between 9 and 11, it will of course cause a
flickering of the preceding digit (330.19 and 330.21).
In such a case, the correct readout is the larger 4-digit
number (330.2) and the fifth digit is acceptably stable.
18. Press [L/Q].
19. Insert as DUT the first inductor specified in
Table 5-2 (check number 10).
20. Verify that the displays are within the ex-
tremes shown for check number 10 in Table
5-2, if the inductor is within the tolerance
specified by Table 5-1.
21. Similarly check the inductor specified in
check number 11.
Table 5-2: Performance Verification