92
1693 RLC Digibridge
Operation
Character
Sequence
Purpose
Allowed
Characters
Meaning
1
Status
(space)
Normal Operation
U
Underrange held (reduced accuracy)*
O
Overrange held (reduced accuracy)*
E
End extension of range 1 or 1 (reduced accuracy)**
Invalid measurement due to signal overload
I
2
Value of delta RLC or
ratio
(space)
Normal display
-
Delta RLC Mode
/
Ratio Mode: Nominal Value/RLC Value
3
Parameter
L
Inductance
C
Capacitance
R
Resistance
G
Conductance
Z
Impedance
Y
Admittance
4
Normal or ratio mode
(space)
Normal mode (one of 3 describe above)
-/
Ratio Mode: RLC Value/Nominal Value
5,6
Units
(space)H
Henries
mH
Millihenries
uF
Microfarads
nF
Nanofarads
(space)S
Siemens
mS
Millisiemens
uS
Microsiements
(space)%
Percent difference from nominal value
(space)Ω
Ohms
kΩ
kiloohms
7
Format
(space)
8
Sign
(space)
Positive Value
-
Negative Value
9...15
Number
0123456789.
(space)
Measured number, right justified in format field, like the
RLC display except the zero before the decimal point is
explicitly provided and this number can be as long as seven
characters. ***
16
(CR)
The standard “carriage return” and “line feed” characters;
end of string.
17
(LF)
Table 3-21: Data output format for RLC value, delta%, deltaRLC, or RATIO
* Underrange generally occurs when the measured impedance is lower than the nominal low limit (for automatic range change)
and there is a lower-impedance range. Overrange generally occurs when the measured impedance is higher than the nominal
high limit (for automatic range change) and there is a higher-impedance range. In either case, an appropriate range change
will improve accuracy. If there is no appropriate range to change to, the status is “range end ex tension “; see below. Range
limits are different if CONSTANT VOLTAGE OR CONSTANT CURRENT has been selected; see paragraph 3.5.6.
** Range 1 end extension if DUT impedance exceeds that of the largest R or L (smallest 0) in basic range 1. Range 4 end
extension if DUT impedance is less than that of the smallest R or L (largest 0) in basic range 4. No possibility of changing range
to improve accuracy.
*** If the measurement is outside of the display range (display goes blank), 9999999 is output in this numeric field.
Содержание RLC Digibridge 1693
Страница 12: ...x Table of Contents 1693 RLC Digibridge This page is intentionally left blank...
Страница 24: ...xxii 1693 RLC Digibridge Abbreviated Specifications This page is intentionally left blank...
Страница 122: ...98 1693 RLC Digibridge Operation Table 3 28 Truth table of compacted binary format for output data...
Страница 130: ...106 1693 RLC Digibridge Theory Figure 4 1 Block diagram of the 1693 RLC Digibridge...
Страница 163: ...139 1693 RLC Digibridge Service and Maintenance Table 5 11 Entry of limits Table 5 12 Bin assignment check...
Страница 171: ......