Diana Peristaltic Pump, ICU Medical
Page 25
Note:
Pressing
“A
ccept
’
will update the calibration of the unit to account for the over/underfill of
the transfer as indicated on the screen (difference between specified volume/weight vs
measured weight).
6.
If the measured weight entered is not compliant with
facility’s g
uidelines, a recalibration may
be performed.
Note:
By pressing Calibrate, the unit will adjust its calibration settings based on the previous
transfer, however, a new calibration process begins to ensure accuracy.
3.4.2.
Warnings during Calibration Process
The following warnings can be encountered during an incorrect calibration process.