Diana Peristaltic Pump, ICU Medical
Page 15
3.
Operating the Diana Peristaltic Pump Module
3.1.
Tubing Set Installation / Removal
The following section describes how the user of the Diana Peristaltic Pump Module properly inserts the
tubing set into the unit.
Note:
Incorrect insertion of the tubing set may damage the unit.
1.
The Diana Peristaltic Pump should be placed with the display view facing the user.
2.
Before opening, inspect the tubing set package. If packaging is not intact, discard it and use a
new set.
3.
Using aseptic technique, open the package and remove the tubing set.
4.
Locate the embossed "arrow" pointing downwards on the handle of the tubing set. Locate a "red
clip" behind the handle.
5.
Insert the tubing set with the "arrow" pointing downward as seen below. Push the red clip
downwards into the recess provided until a "snap-
in” sound
is heard.