The first warning message is displayed when a discrepancy between expected and measured
weight is more than 30%.
Pressing the “OK”
returns user to Prime part of the workflow. Repeat
the priming and calibration to ensure the accuracy of the unit.
The second warning message is displayed when a discrepancy between expected and measured
weight is within 10-30% range. Pressing "Yes" button confirms the user's entry and completes
the calibration process. Pressing "No" button returns user to Prime part of the workflow.
3.5.
Basic Set Up and Transfer
Upon proper calibration, the unit allows a desired volume to be transferred and the speed to be
set up before initiating a transfer.
1.
Enter Volume to be transferred and select the speed. Press
“
Next
”
to confirm the transfer
process.
Wrong Calibration Alert
Calibration Out of Range