Diana Peristaltic Pump, ICU Medical
Page 27
Note:
Pressing
“
Next
”
without entering a volume will enable the
“
Infinite Transfer
”
mode. Pump
will indicate that “Infinite Transfer” mode is about to be initiated, volume number will change to
an infinity symbol. Infinite mode will continue until the user manually stops the transfer.
2.
Unit will move to the Compound tab screen. After confirming that entered information is
accurate, press
“
Start
”
to initiate transfer. The unit will begin fluid transfer and the progress
bar will indicate the status of the transfer. The volumetric measurement will provide an
indication of what has been transferred into the bag. Pressing Stop will interrupt the
transfer.
3.
Once the transfer is complete the system displays the Compounding Completed message.
The system will automatically return to the Set Up screen (the duration of Compounding
Completed message display is configurable in the Advanced Settings).
4.
Disconnect the Tubing Set and place the destination container on the External Scale for
weighing. Check the weight and verify that the correct volume is transferred.
5.
After compounding is complete follow
facility’s
medication handling protocol and
procedures.