Page 24
Note:
The speed will be based on the volume and viscosity of the fluid. To adjust the speed of the
pump, tap the arrows on either side.
3.
Tare the external scale and press
“S
tart
”
to initiate the transfer.
Note:
To ensure proper calibration, the destination container must be detached from the tubing
set during the Weighing Process. Failure to do so will results in inaccurate calibrations. Ensure
destination container is reconnected to the tubing set before pressing
“
Start
”
.
4.
Enter the weight measured by an external scale into the screen as seen below. If the measured
weight is acceptable, press
“
Accept
”
.
5. After a successful calibration, the unit will display a confirmation that calibration has been
completed and the user will be directed to the Set Up menu.