Press the “Next" button. The next step is to start the
priming and calibration process.
3.3.1.
Changing The Specific Gravity
The specific gravity of a drug can be changed by the user on the drug menu screen. Changes to
this option require a user permission to be enabled by the authorized user in the "Advanced
Settings."
Note:
The default specific gravity for most drugs is set to 1.0 g.
1.
If user selects a drug from the database, the default specific gravity is set to 1.0 g. User will
see a blue square with a yellow triangle containing an exclamation mark (see below), this
means that the specific gravity can be changed by the user to a different value. When user
clicks on the specific gravity field a warning message will appear advising that user is about to
change the specific gravity for that particular drug (see below).
Note:
If user presses
“
Cancel
”
, system will return to the Drug Set up screen with the specific
gravity selected at the default of 1.0 g.
2.
Press "Continue" to change the specific gravity of the selected drug by manually entering using
the keypad. Then press the "Enter" button.