Electrical Operating Power Range 100
–
240 V AC
0.8
–
0.4 A
50
–
60 Hz
Electrical Fuse
5.3.
Lithium-Ion Coin Battery Replacement
Peristaltic Pump relies on a 3-Volt Lithium battery (
CR2032)
. Service on the battery is
recommended to be done only by an authorized ICU Medical representative. Incorrect
insertion of the Lithium-ion battery may result in a damage of the unit.
Battery location can be identified by this symbol inside the unit.
1.
Make sure the unit is powered-off and is unplugged.
2.
Open the unit and locate the above symbol where the lithium-ion battery holder is located.
3.
Remove the old battery if any.
4.
Insert the new battery into the unit with positive (+ Symbol) side facing up.
Note:
Incorrect insertion will damage the unit.
5.4.
Fuse Replacement
Peristaltic Pump uses two fuses located in the rear of the unit; fuses are a part of the power
inlet module. Service on the fuses can be done by the end-user (end-user buy the fuses from
local hardware/electrical store). Incorrect insertion of the fuses may result in no power to the
unit.
1.
Make sure the unit is powered-off and power cord is unplugged.
2.
Using flat head screw driver remove the fuse cap from power inlet module, by turning
the fuse holder cap in a counterclockwise direction.