38
Instruction Manual C200
4.5 Automatic zero adjustment of gauges
-
Place the master in the measuring device
-
Use the encoder to select and start the
“CALIBRATION”
menu.
The calibration menu is password protected, if the numeric display indicates
“PASS.Cd.”
( See
BASIC SETUP “LA – PASS.Cd.”
)
-
The numeric display alternately indicates
“CAL. 1”
and the currently
measured value
.
-
Push the encoder button to execute the automatic zero adjustment. The master value is adopted
as measurement value by the gauge and the column gauge returns to the measuring mode.
The automatic zero adjustment can also be executed in the measuring mode by actuating the encoder
button, foot or hand switch, or an external contact. The
“BASIC SETUP”
( menus
L2
…
L5
) must be
programmed accordingly in order to allow this option. The automatic zero adjustment can also be
triggered by a timer forced mode. This function is programmed in the menu
L9
of the
“BASIC SETUP”
.
4.6 Automatic gauge calibration
The automatic gauge calibration is activated by programming two masters.
Procedure :
-
Place one of the two masters in the measuring device
( or insert the gauge head into one of the masters ).
-
Use the encoder to select and start the
“CALIBRATION”
menu.
The calibration menu is password protected, if the numeric display indicates
“PASS.Cd.”
( See
BASIC SETUP “LA – PASS.Cd.”
)
-
The numeric display alternately indicates
“CAL. 1”
and the currently
measured value
.
-
Push the encoder button to confirm the first master.
-
The numeric display alternately indicates
“CAL. 2”
and the currently
measured value
.
-
Place the second master in the measuring device ( or insert the gauge head into the second master ).
-
Push the encoder button to confirm the second master.
-
Now, the column gauge automatically calibrates the currently selected gauge. The second master
value is adopted as measurement value by the gauge and the column gauge returns to the measuring
mode.
The automatic gauge calibration can also be executed in the measuring mode by actuating the encoder
button, foot or hand switch, or an external contact. The
“BASIC SETUP”
( menus
L2
…
L5
) must be
programmed accordingly in order to allow this option.
The automatic gauge calibration can also be
triggered by a timer forced mode. This function is programmed in the menu
L9
of the
“BASIC SETUP”
.
4.7 Multi gauging ( C1 … C8 )
The
C200
column gauge allows programming up to 8 independent gauges (
C1
to
C8
).
The unit, resolution, nominal size, tolerance, measuring mode and master values can be programmed for
each individual gauge. You can switch over to the individual gauges (
C1
to
C8
) in the “
MEASURING”
mode by pressing and turning the encoder at the same time, if the function “manual gauge selection” is
activated. The gauge selection can also be performed in the measuring mode by actuating the encoder
button, foot or hand switch, or an external contact. The
“BASIC SETUP”
( menus
L2
…
L5
) must be
programmed accordingly in order to allow this option.
The gauge selection (
C1
to
C8
) is performed automatically, if
“Aut. on”
was set in the
“BASIC SETUP”
menu (
“L1
–
Auto.rE.”
). A change in the measured value of a gauge (
C1
to
C8
) switches over the
column automatically to the corresponding gauge.
Note : This operating mode is preferred mainly for applications with several bore gauges.