Instruction Manual C200
GrAdE
The grading mode can be activated or deactivated independently for each gauge (
C1
to
C8
)
in this menu. The number of grading groups ( classes ) can be set from
1
to
30
and deter-
mines in how many, equal fields
( linear pitch )
the tolerance range is split.
Note : If no tolerance limits have been specified, the numeric display in the measuring
mode indicates “
------ ”
.
t.F.-CAL
In this menu the timer forced calibration can be activated for each gauge (
C1
to
C8
). At first
the time interval is set to a value from 15 minutes up to 63 hours and 45 minutes ( in 15 min-
ute steps ). Then those gauges are selected for which the timer forced calibration shall be ac-
tivated. The same time interval applies to all selected gauges. The column gauge automati-
cally switches to the calibration mode, as soon as the time interval has elapsed.
The
Calibration
- LED flashes and the measuring mode is locked. After pushing the encoder
button, a calibration of the displayed gauge can be performed as described at 3.4.1.
Note : If the timer forced calibration has been activated for several gauges, then those
gauges that were not selected for display on the column gauge as the time interval
elapsed, must be calibrated as soon as they are selected for display.