Instruction Manual C200
Example 3 : Connection of pneumatic gauge heads
Example 4 : Connection of different probes, gauges, sensors and measuring instruments
Oil and dirt can destroy
the measurement system !!!
For more information see
IMB-ae1
Manual,
Doc. D2MF122 081
The example depicts a
C200
configuration for connection of
2 inductive probes (
IMB-im2
),
2 incremental probes with
1Vss output (
IMB-dm2
), one
pneumatic plug gauge (
IMB-ae1
),
2 Mitutoyo dial gauges (
IMB-mi2
)
and one calliper gauge ( Sylvac, Tesa,
Mahr, etc. ) with Opto RS232 output
(
IMB-sm1
).
The column gauge allows the connection
of up to 8 pneumatic measuring converters
IMB-ae1
. An 8-by-6 mm hose connector
( 8 mm -> outer diameter; 6 mm -> inner
diameter ) is used to establish the com-
pressed air supply. A hose piece with a
length of 35 mm is required in order to
pneumatically interconnect two
IMB-ae1
modules. Use dry, oil-free and filtered
( 5 µm ) compressed air only, ranging from
2.6 … 2.8 bars ( 39 … 42 psi ) to connect
to the
IMB-ae1
measuring converters
( recommended air pressure ).