Instruction Manual C200
In the Basic Setup menus
L2
to
L5
controlling of the
C200
via the encoder button and the foot or hand
switch connections
Ft1, Ft2
as well as the IMB foot switch
Ft3
is set. The encoder button and each one of
the 3 foot or hand switch connections can be assigned any function from the list.
Description of the different functions :
oFF *
- No function assigned to the encoder button or the respective foot or hand switch.
GAGESL.
- If this function has been assigned to a particular button or switch, and if that button or switch
is then activated in the measuring mode, you will automatically switch to the next activated
gauge from
C1
to
C8
.
CALibr.
- If this function has been assigned to a particular button or switch, and if that button or switch
is then activated in the measuring mode, an automatic zero adjustment or an automatic
gauge calibration will be performed on the currently active gauge (
C1
to
C8
).
dyn.con.
- If this function has been assigned to a particular button or switch, and if that button or switch
is then activated in the measuring mode, the dynamic measurement function can be
controlled.
There are two control options :
EdGe
:
1. Actuation starts the dynamic measurement function
2. Actuation stops the dynamic measurement function
StAtE
:
The dynamic measurement function runs for as long as the button or switch is
activated.
dYn.diS.
- If this function has been assigned to a particular button or switch, and if that button or
switch is activated in the measuring mode, you will switch between the dynamic modes
(
Min, Max, TIR, etc.
) with display of the measured values. The currently active gauge
however, must have been programmed for a dynamic measurement mode. This function
enables viewing the four results for the
Min, Max, TIR,
and
Mean Value
of the component
after a dynamic measurement has been taken.