Instruction Manual C200
3.8 Error messages / Corrective actions
The numeric display indicates
Operating and Programming Errors
caused by the user as well as
System Errors
of the column gauge.
3.8.1 Operating and programming errors
Err. xx :
error message ( Err ), error number ( xx )
Error
Error description
Corrective action
Err. 1
All measuring inputs are switched off in the
probe interconnection or one of the measuring
inputs is defective.
1. Select
MEASURING INPUTS
and check to
make sure that at least one measuring
input is activated.
2. Select
CALIBRATION
and then
AdJuSt
and
check the individual measuring inputs.
Err. 2
The wrong password was entered.
Enter the correct password.
(
If you have forgotten the password, you can use the default
password
200879
. )
Err. 8
Automatic gauge selection is activated.
One measuring input has been assigned to
several gauges.
The automatic selection of an active gauge is
not possible because this input also changes
the measured values of various other gauges.
There are two ways to eliminate the error :
1. Select
BASIC SETUP – L1
menu and set
Aut.oFF
.
2. Select
MEASURING INPUTS
and deactivate
the input that was assigned repeatedly.
Err. 10
For automatic gauge calibration, the same
master was placed in the measuring device
twice.
Repeat the calibration procedure using two
different masters.
Err. 11
The discrepancy between the two masters is
too large. ( discrepancy > 6.5 mm )
Repeat the calibration procedure using two
masters that are not that different.
Err. 12
The discrepancy between the two
programmed master values is too large.
( discrepancy > 6.5 mm )
Select the menu
MASTER VALUE
and
reprogramme the master values.
Err. 13
Error performing the automatic gauge
calibration.
The calibration factor is too small.
The difference between the programmed master
values is much smaller than the difference
between the measured master values.
Repeat the calibration procedure and check the
master values programmed in the
MASTER
VALUE
menu.
Err. 14
Error performing the automatic gauge
calibration.
The calibration factor is too large.
The difference between the programmed master
values is much larger than the difference
between the measured master values.
Repeat the calibration procedure and check the
master values programmed in the
MASTER
VALUE
menu.