Page 45
Preserve system
CAUTION!
Material damage risk
Preservation is used for protection of the membrane against
microbiological contamination. No later than 3 months after a
preservation, the system must be taken into operation (see hint
with respect to rinsing in chapter 7, „Initial start-up“) or must be
preserved again.
PLEASE NOTE
For the next steps, do not remove the Personal protective equip-
ment (gloves, protective glasses) used when preparing the pre-
servation agent.
1
Cover the transportation box interior with rugs or similar material and
protect the system parts within the fine filter surrounding, since the
preservation dilution may spill over.
2
Pour preservation dilution into filter housing cover.
3
Remount filter housing cover.
4
Connect raw water supply hose with system separator to raw water
supply.
5
Check mains cable for damage and insert mains plug into socket.
The green LED flashes.
6
Open raw water supply.
The pressure switch (permeate) is activated.
The systems starts up and production commences.
The green LED is lit.
7
•
Venting: Push red button on fine filter very shortly until water
exits
•
Rinse-in preservation dilution for a period of 20 secs
8
•
Pull mains plug
•
Cut raw water supply and disconnect
•
Disconnect concentrate hose
9
Let hoses run empty. Disconnect hoses from transportation box con-
nections and stow away within transportation box.
10
Mark system clearly as „Preserved“.
Содержание WL-RO-170
Страница 1: ...Reverse osmosis plants WL ROB 170 WL RO 170 WLROB170 ENx WLROB170 EN Manual...
Страница 21: ...Page 21 5 5 Dimension sheet Fig 3 Front Top view Permeate Connection R3 4...
Страница 22: ...Page 22 Fig 4 Lateral view Fig 5 Front view of securing and fixing plate Concentrate Hose 8 Wastewater drain...
Страница 46: ...Seite 46 This page intentionally left blank...