Verifying the Probe Zero at a Later Time
1) Press the
MENU
key once to activate the menu items tab. Press the
MENU
key multiple times to tab right, and the
ESC
key multiple times to
tab left, until the
DATA
menu is highlighted and displaying the submenu
items.
2) Use the
UP
and
DOWN
arrow keys to scroll through the sub menu
items until
OPEN
is highlighted.
3) Press the
ENTER
key to display the
Open Group List Box.
4) Use the
UP
and
DOWN
arrow keys to scroll through the list menu items
until the
group name containing the probe zero data
is highlighted.
5) Press the
ENTER
key to load the zero data group into memory, followed
by pressing the
OK
key to confirm loading the group into memory.
Note:
The user will always use this group for the zero verification
procedure, as it contains the original settings and measurement data of the
probe zero calibration standards.
6) Apply a drop of couplant to the cal standard or transducer, and attach it
to one end of the cal standard. Rotate the transducer clockwise and