hydratight
150
Chapter 13 Additional Features
13.1 Pulse Width
The
BOLTSCOPE PRO
is equipped with an adjustable pulse width option. Pulse
width; refers to the duration of time the pulser is left on. This time results in
increased energy sent into the test material. There are three width options
(
SPIKE, THIN,
and
WIDE
). The
SPIKE
setting may be desirable for high
resolution and general applications to decrease the overall noise. This can be
considered the normal or standard setting. When additional energy is needed,
more penetration, the
THIN
and
WIDE
options may be necessary. The
procedure to change the pulse width is outlined below:
Selecting the Pulse Width
1) Press the
MENU
key once to activate the menu items tab. Press the
MENU
key multiple times to tab right and the
ESC
key multiple times to
tab left until the
TUNE
menu is highlighted and displaying the submenu
items.
2)
Use the
UP
and
DOWN
arrow keys to scroll through the sub menu items
until
PULSE WIDTH
is highlighted.
3)
Use the
LEFT
and
RIGHT
arrow keys to scroll the
Pulse
options.
4)
Once the correct
Pulse
is displayed, press the
MEAS
key to return to
measurement screen.
13.2 Alarm Mode
The
Alarm Mode
feature of the
BOLTSCOPE PRO
provides the user with a
method of setting tolerances, or limits, for a particular application requirement.