BOLTSCOPE PRO
- Bolt Tension Monitor
125
Rectified View
Rectified
The rectified A-Scan mode shows half of the waveform. Either the positive or
negative peaks are displayed, depending on the polarity selected.
It’s important to note that a measurement must fall inside the viewable display
range in order to see the waveform. However, even if the waveform is outside of
the viewable area, a measurement can be taken and viewed on the digital
readout. If the waveform is outside of the display range, you can manually
change the range by adjusting the
Delay
and
Width
values, or you can use the
Auto Find
feature located in the
UTIL
menu in the tabbed menu items. Refer to
page
Error! Bookmark not defined.
for additional information on the
Auto Find
feature.
The following is a list of the viewable features on the display:
A)
Stability of Reading Indicator
– Indicates the stability of the return echo on
a scale of 1 to 6 – the solid bars displayed in the figure above indicate a
repeatable signal. If the
BOLTSCOPE PRO
is displaying a reading from
memory, the repeatability indicator will be replaced by the text
“MEM”
.
B)
Battery life indicator
– Fully charged batteries will appear filled in solid.
Note: The diagram shows the batteries at approximately 50%.
C)
Length Measurement
– Digital readout of quantity (english or metric).
D)
Detect Indicator
– The broken vertical line displays the zero crossing (flank)
detection point on the waveform where the measurement has been obtained.
Notice that the digital quantity readout is the same as the location of the
detect indicator according to the measurements labels F, also in the diagram.