BOLTSCOPE PRO
- Bolt Tension Monitor
55
Storing the Reference Length
1) Press the
MEAS
key once to activate measure menu items. Press the
MEAS
key multiple times to move right and the
ESC
key multiple times
to move left until the
LOC
cell is highlighted.
2) Press the
ENTER
key to display the
Group View Box
.
3) Press the
UP
,
DOWN
,
LEFT
, and
RIGHT
arrow keys to scroll the target
cell cursor to the desired storage location.
Note:
The first column must be used to store reference lengths.
4) Press the
ENTER
key to save the current reading in the highlighted cell
location.
5) Abort the
Grid View Box
by pressing the
MEAS
key at any time.
6) Proceed to the next section to perform and store the first
Probe Zero
verification. DO NOT REMOVE THE TRANSDUCER FROM THE
BOLT.
The next two sections outline the procedures of how to perform a probe zero
calibration by calculating a zero value, or alternatively adjusting the temperature
value to match the existing probe zero value. When should I use the
Calc Zero
feature versus adjusting the
Temperature
, and why?
The only time the
Calc Zero
feature should be used is when a different
BOLTSCOPE PRO
, transducer, or transducer cable is being used, other than the