BOLTSCOPE PRO
- Bolt Tension Monitor
73
Measuring Calibration Standards – For the First Time
1) With the menu items still active from the previous section, press the
MENU
key multiple times to tab right and the
ESC
key multiple times to
tab left until the
CAL
menu is highlighted and displaying the submenu
items.
2) Use the
UP
and
DOWN
arrow keys to scroll through the sub menu items
until
ZERO MODE
is highlighted.
3) Use the
LEFT
and
RIGHT
arrow keys to scroll the zero mode options
until the
TWO POINT
option is displayed.
Note:
The two-point option allows the user to use two separate points for
the probe zero calibration. Note: The one point option can optionally be
selected if only one cal bar will be used. Although the procedures below
are described with the two-point option in mind, they will also work in the
case of a one-point calibration.
4) Use the
UP
and
DOWN
arrow keys to scroll through the sub menu items
until
MEASURE ZERO
is highlighted.
5) Press the
ENTER
key to display the
Two Point Zero
list box.
6) Press the
UP
and
DOWN
arrow keys to scroll through the
Two Point
Zero
list until
PHY LEN 1
is highlighted.