15) Use the
UP
and
DOWN
arrow keys to scroll through the sub menu
items until
MEASURE ZERO
is highlighted.
16) Press the
ENTER
key to display the
Two Point Zero
list box.
17) Press the
UP
and
DOWN
arrow keys to scroll through the
Two Point
Zero
list until
ULTRASONIC 1
is highlighted – 3” bar or 1” side of glass
block.
18) Press the
ENTER
key to store the measured value of the 3” bar or 1”
side of glass block.
19) Apply a drop of couplant to the bolt or transducer, and attach it to one
end of the
6” calibration bar
(or 3” side of glass block). Rotate the
transducer clockwise and counter clockwise applying a small amount of
pressure to eliminate any excess couplant between the transducer and
cal bar surface. Remove and replace the transducer a couple of times
in the exact same location on the cal bar, rotating it each time to obtain
the minimum value measured.
Be sure to always place the transducer
in the same location
.
This will help to eliminate any potential
measurement errors caused by changing the sound path.
20) Press the
UP
and
DOWN
arrow keys to scroll through the
Two Point
Zero
list until
ULTRASONIC 2
is highlighted – 6” bar or 3” side of glass
block.
21) Press the
ENTER
key to store the measured value of the 6” bar or 3”
side of glass block.
22) Press the
UP
and
DOWN
arrow keys to scroll through the
Two Point
Zero
list until
CALC ZERO
is highlighted.
23) Press the
ENTER
key to calculate the zero.