hydratight
74
Note:
PHY LEN 1
is the actual physical length of the first calibration
standard. If the glass block is being used, the 1” side should be used for
PHY LEN 1
.
7) Press the
ENTER
key to display the
Digits Edit Box
.
8) Press the
UP
and
DOWN
arrow keys to scroll the highlighted value.
9) Press the
LEFT
and
RIGHT
arrow keys to scroll the digit locations.
10) Repeat steps
8 & 9
until the
PHY LEN 1
value is correctly displayed.
11) Press the
OK
key to return to the
Two Point Zero
list, or
ESC
to cancel
entering the
PHY LEN 1
.
12) Press the
UP
and
DOWN
arrow keys to scroll through the
Two Point
Zero
list until
PHY LEN 2
is highlighted.
Note:
PHY LEN 2
is the actual physical length of the second calibration
standard. If the glass block is being used, the 3” side should be used for
PHY LEN 2
.
13) Press the
ENTER
key to display the
Digits Edit Box
.
14) Press the
UP
and
DOWN
arrow keys to scroll the highlighted value.
15) Press the
LEFT
and
RIGHT
arrow keys to scroll the digit locations.
16) Repeat steps
14 & 15
until the
PHY LEN 2
value is correctly displayed.
17) Press the
OK
key to return to the
Two Point Zero
list, or
ESC
to cancel
entering the
PHY LEN 2
.
18) Press the
UP
and
DOWN
arrow keys to scroll through the
Two Point
Zero
list until
VELOCITY
is highlighted.
Note:
This refers the velocity of the probe zero cal bars. Since there is a
velocity for each cal bar, use the velocity of the longest cal bar as it will be