Sonicaid FM800 Encore Series Instructions for Use
94
12
11B
Sonicaid TREND
12.1
98B
Introduction
Sonicaid TREND is a software option available with all FM800E series monitors. It
measures fetal heart rate parameters at regular intervals, and describes the trace in a way
that is quantitative and not qualitative. TREND is not intended as a replacement for skilled
visual interpretation of the trace, but it does help long-term changes in the fetal heart rate
pattern to be assessed.
Caution: TREND is valid only during the first stage of
labour.
The trace trend data provided by the Sonicaid TREND is
intended to assist, not to replace, the physician's visual
assessment of a trace.
No guidelines on interpretation or limits of normality are provided, but the clinician can use
the numeric values to identify and quantify the relative changes in fetal heart rate
parameters over a period of time.
Numerical description of the trace allows direct comparison between traces. It also
provides training support for trace interpretation, and readily available data for clinical
research projects.
235B
IMPORTANT
TREND provides numeric & objective trending information on key FHR trace parameters
only. It does not provide any form of trace interpretation
. Interpretation and diagnosis of
the record remain the responsibility of the appropriately qualified medical staff.
Note
: Sonicaid TREND is not approved for sale in the USA and Canada.