Sonicaid FM800 Encore Series Instructions for Use
11
1.7
25B
Multiple Portable Socket Outlets
(including isolation transformers)
Note
: an isolation transformer is a particular kind of multiple socket outlet.
It is not recommended to power a medical system from a multiple portable socket outlet
which is not supplied from an isolation transformer (IEC601-1-1/EN60601-1-1
Amendment 1).
If such an outlet is in use, it should comply with the requirements of Annex EEE.2 of
IEC601-1-1/EN60601-1-1 Amendment 1.
Do not exceed the power rating for the multiple portable
socket outlet.
Do not place multiple portable socket-outlets on the floor.
This is to protect against mechanical damage and the
ingress of liquids.
Multiple portable socket-outlets supplied with the system
must not be used for powering equipment which does not
form part of the system. This is to prevent increased leakage
currents, and overload of the multiple portable socket outlet.
If the system has been specified for use with an isolation
transformer, do not connect any non-medical electrical
equipment which forms part of the system directly to the
wall outlet. This is to prevent excessive leakage currents
Non-medical electrical equipment situated in the patient
environment (within 1.5 metres of the patient) must be
powered via an isolation transformer, to limit leakage
current.
For more information on the connection and use of isolation transformers, consult the
user manual for the medical system you have purchased.