Sonicaid FM800 Encore Series Instructions for Use
35
The default is for the paper-out buffer to be enabled (i.e. FM800E will store data if the
paper runs out). To switch off the paper-out buffer, or switch it on again after it has been
switched off:
> SETUP > NEXT PAGE > SYSTEM SETTINGS > access code (2755)
> PRINTER.
> PAPER OUT BUFFER.
175B
Paper Speed
To change the paper speed:
> SETUP > NEXT PAGE > SYSTEM SETTINGS > access code (2755)
> PRINTER.
> PAPER SPEED once or twice, as required.
176B
FHR graticule (Sonicaid paper only)
The FHR graticule can be printed at intervals of 5 bpm or 10 bpm.
To alter the FHR graticule:
> SETUP > NEXT PAGE > SYSTEM SETTINGS > access code (2755)
> PRINTER.
> FHR GRATICULE.
177B
FHR scale for twins (Sonicaid paper only)
If twins are being monitored, each FHR trace can be printed on its own scale, or both on
the same scale.
To alter the scale for twins:
> SETUP > NEXT PAGE > SYSTEM SETTINGS > access code (2755)
> PRINTER.
> TWIN FHR.
TWIN SCALES setting is not recommended for intrapartum
monitoring as the FHR values may exceed the printed scales.
178B
FHR vertical scale
Warning: Where other makes of paper are used, this setting
MUST match the pre-printed grid scaling. Failure to correctly
set this will result in mis-leading traces.
The FHR vertical scale can be 20 bpm/cm or 30 bpm/cm. At 20 bpm/cm the range is
50-210 bpm. At 30 bpm/cm it is 30–240 bpm.
To change the vertical scale:
> SETUP > NEXT PAGE > SYSTEM SETTINGS > access code (2755)
> PRINTER.
> FHR SCALE.