Sonicaid FM800 Encore Series Instructions for Use
93
11.4
97B
Display Buffer
The display buffer is the internal memory where the trace information is stored. The length
of the buffer is such that it can store:
24 hours of data at 3cm/min.
36 hours at 2cm/min.
72 hours at 1cm/min.
Beyond this, the oldest data is discarded and the buffer then retains the latest 24, 36 or 72
hours of trace information.
The information stored in the Display Buffer can be viewed
using the scroll buttons (See 11.3), but it cannot be
extracted or printed.
The display buffer is cleared if the unit is turned off, or in the
event of a power failure. Any trace information stored will be
lost.