14.2
110B
Using FM800E with Sonicaid Centrale, or other
CRS systems
255B
Connecting FM800E to Sonicaid Centrale or other CRS systems
There are a number of ways in which FM800E may be connected to Sonicaid Centrale or
other CRS’s. Some of the combinations are described below but the list is not exhaustive.
For other ways of connecting consult the Central Review System administrator.
1 Directly - using RS232
Connect an RS232 lead to the RS232 CRS connector on the rear of FM800E. See
Section 2.4. Connect the other end of the lead to a spare RS232 input port of the
Central Review System.
2 Indirectly – using LAN
For full details of PC connections, and instructions for using the system, see the
documentation supplied with the Central Review System.
At switch on, the monitor sends an ID packet to the CRS. If
the FM800E is not connected to the CRS at power on, the
CRS may not register the presence of the FM800E. In such
cases, ALWAYS connect the FM800E to the CRS before
applying power to the FM800E.
Setting up FM800E for use with Sonicaid Centrale, Fetalcare, Obstetric
Archive or other CRS
1 > SETUP > NEXT PAGE > SYSTEM SETTINGS > access code (2755).
2 > SERIAL INTERFACE.
3 Make sure the RS232 CRS button says ‘HP’. If ‘Sonicaid’ is displayed, then press the
RS232 CRS button to change to ‘HP’.
4 > EXIT.
5 Switch off FM800E. Wait for about 10 seconds.
6 Switch FM800E on again.