Sonicaid FM800 Encore Series Instructions for Use
112
18
17B18B
Maintenance
WARNING
It is very important that all instructions in the Maintenance
section are followed carefully.
18.1
124B
User maintenance
The checks below can be performed by any user of the equipment.
256B
Mechanical inspection
Every three months:
1 Inspect the AC supply cable, transducers, and all other assemblies and connectors for
loose or broken parts, or any other damage.
2 Pay particular attention to the AC supply socket.
3 Look carefully for cracks which may allow the ingress of liquids or gels.
4 Replace any broken or damaged transducers or cables.
5 If there is damage to the main FM800E unit, contact your local Huntleigh Healthcare
Ltd representative.
257B
Cleaning the print head on the printer
1 Pull the paper tray out as far as it will go.
2 Remove the paper pack.
3 Using a lint-free cloth and pure alcohol, wipe along the full width of the print head,
which is beneath the plastic edge of the paper compartment.
4 Replace the paper tray and paper pack.
258B
Check NBP cuffs and hose
Once a month:
1 Check the NBP hose. Straighten out any kinks and distortions.
2 Check the cuff(s) for wear and damage.