HumaStar 600
| User manual
76
For operation, proceed as follows:
1. Define a Twin set. Press the
Twin definition
tab:
- Define a name which defines a given set of controls.
- Select the prefix for the high and low (or normal/abnormal) levels. It is im-
portant that they must coincide with the Control Vial Id (given in the Control
Set).
2. Link the samples. Press the
Link Samples
tab:
- Select the Twin Id.
- Press the
Refresh Lists
button to display all controls which correspond to
high and low level.
- Select for each level a control and press
Link
.
FIGURE 76
Coupled quality control is a
useful tool if both controls of
each pair are taken in similar
conditions. Observe index num-
ber, date, time and lot number
and verify that data were taken in
close conditions.
FIGURE 77