No new sample entries are allowed in this period. A warning is issued if cooling
system is off or defective.
Automatic cycle
includes reagent testing, integrity check, test analysis and cal-
culations. Also printouts can be generated in this period.
Post-automatic
includes cuvette washing, probe cleaning and conditioning and
remaining printout.
If “Non Stop operation” has been selected, instrument remains idle until new
samples are introduced or the check box gets deselected.
4.1.1 CLOT DETECTOR
System counts with two clot detectors, one in each probe. They operate based
on a differential pressure principle.
Clot detector is installed in
Maintenance > Service > Parameters > Instrumentals > Others
Detectors automatically calibrated when automatic cycle starts. Several work-
ing conditions can be adjusted. They are defined in section 3.2.6.1:
Maintenance > Parameters > General
and modified with Supervisor privileges.
Operation of clot detectors can be enabled or disabled; if clot is detected, then
sample can be repeated or not.
The clot condition will be posted in the details of the sample result.
If sample is discarded, it will continue as in process and a message will be shown
in the window of messages in main menu. Clot detector version is posted in the
ErrorsLog.txt file.
Содержание HumaStar 600
Страница 1: ...HumaStar 600 User Manual Cat No 16660 001...
Страница 2: ......
Страница 4: ......
Страница 14: ...HumaStar 600 User manual 10...
Страница 44: ...HumaStar 600 User manual 40...
Страница 107: ...ISE Module Cat no 16663 03 103 7 ISE MODULE CAT NO 16663 03 See HumaStar 600 ISE MODULE User Manual 16663 1...
Страница 108: ...HumaStar 600 User manual 104...
Страница 129: ...125...
Страница 130: ...HumaStar 600 User manual 126...
Страница 131: ......