6.1.1 ENDPOINT
6.1.1.1 Single point end point
Readings
A single absorbance reading (A
1
) is taken at specified time after reagent addition.
Other absorbance reading (Ae
1
) may be taken immediately after the first read-
ing for extra precision.
Measurement
M = A
1
- B.
where B is the measurement of the reagent blank if required, otherwise set to 0.
The reagent blank determination is analogous to sample reaction.
If extra precision is required, measurement is computed as the average of the
first and extra precision readings as
M = ((A
1
+ Ae
1
) / 2) - B
Limitations
Reagent times for R2/R3 (if any) shall be 0.
6.1.1.2 Two point end point
Readings
The first absorbance reading (A
1
) is taken just before last reagent addition. The
second absorbance reading (A
2
) is taken at specified time after last reagent ad-
dition.
Measurement
The measurement is calculated as
M= (A
2
-F*A
1
) – B
Where F is a
Volume Factor Correction
given by
F= (V1 + Vs ) / (V1 + V2 + Vs)
With V
1
: first reagent volume
V
2
: second reagent volume
V
s
: sample volume
Содержание HumaStar 600
Страница 1: ...HumaStar 600 User Manual Cat No 16660 001...
Страница 2: ......
Страница 4: ......
Страница 14: ...HumaStar 600 User manual 10...
Страница 44: ...HumaStar 600 User manual 40...
Страница 107: ...ISE Module Cat no 16663 03 103 7 ISE MODULE CAT NO 16663 03 See HumaStar 600 ISE MODULE User Manual 16663 1...
Страница 108: ...HumaStar 600 User manual 104...
Страница 129: ...125...
Страница 130: ...HumaStar 600 User manual 126...
Страница 131: ......