Select the method from the list or type in the method Id. Select the calibrator
vial number, concentration and the number of replicates and press
Add
. It is
recommended to use at least 2 replicates for each calibration point.
Press
Ok
when done or
Cancel
to abort. For a multi calibrator repeat this steps
for each test/method.
To remove a method from the calibrator set, select the test and press
Delete Test
.
Once no further modifications are required for the calibrator set, press
Ok
to
finish or
Cancel
to abort.
Once tests are loaded in the grid, concentration values can be edited at all times.
5.4.1.2 Removing a calibrator set
To remove a calibrator set from the list, select the calibrator set from the list
on the right and press
Delete
. Press
Yes
to confirm or
No
to abort.
5.4.1.3 Automatic Dilutions
It is possible to dilute a single calibrator for a multi point calibration. When
A
dd Test
is pressed, the following screen will show up:
FIGURE 59