9 TROUBLESHOOTING
Instrument related problems can be classified into three groups:
1. Operation malfunctions with visual, acoustic or printed warnings.
2. Visible faults or problems.
3. Measurement inconsistencies (for example: GOT method with high disper-
sion).
9.1 Messages and Warnings
Self-explanatory messages are not included in the present listing.
Message
Cause
Action
Change reaction
cuvettes.
All 160 cuvettes are dirty. Replace reaction cuvettes.
Front / Back syringe
exceeded.
Preset limit is surpassed.
Replace at the earliest op-
portunity.
Front / Back pump
turns exceeded
Preset limit is surpassed.
Replace at the earliest op-
portunity.
There is no enough
cleaning solution.
Cleaning or rinsing
solutions are missing.
Replace required System
Wash Solution.
9.2 Visible faults
9.2.1 GENERAL FAULTS
Symptom
Corrective Action
Drops on probe tip after dis-
pensing
Verify hydraulic system in accordance to user’s
manual.
Clean probe tip by submerging in Solution 1 for
5 minutes.
Drops on tip after wash cycle. Verify hydraulic system for leaks or obstructions.
Abnormal noises.
Defective fans.
Moving parts blocked or frozen.
Contact Technical Support.
TABLE 20
TABLE 21
Содержание HumaStar 600
Страница 1: ...HumaStar 600 User Manual Cat No 16660 001...
Страница 2: ......
Страница 4: ......
Страница 14: ...HumaStar 600 User manual 10...
Страница 44: ...HumaStar 600 User manual 40...
Страница 107: ...ISE Module Cat no 16663 03 103 7 ISE MODULE CAT NO 16663 03 See HumaStar 600 ISE MODULE User Manual 16663 1...
Страница 108: ...HumaStar 600 User manual 104...
Страница 129: ...125...
Страница 130: ...HumaStar 600 User manual 126...
Страница 131: ......