HumaStar 600
| User manual
18
BCR
1. BCR: Barcode reading can be activated for sectors, samples and reagents by
checking the corresponding box. When sectors are provided with bar code
identification, it is not necessary to define a number for sector loading.
2. Reagent configuration: Select setting according to your requirements. This
selector defines options for Method, bottle type, expiration date format and
starting position. In case of closed parameters, values are predefined and no
intervention is necessary.
3. Sample configuration: If Id position is not selected, all barcode digits are read.
4. Reagent no read behaviour: If reagent barcode is activated two options are
available for barcode reading error messages. The message can show up
each time an error accures or at the end of the reagent barcode reading.
5. Lot Number: To use multiple lots for the same reagent select the box “Mix
lots”.
Press
Apply Changes
in order to save the changes.
FIGURE 6